Table of contents First page 69 71 Last page
◊ Examine the maintenance records for proper housekeeping. ◊ Is the backwash regime functioning properly? ◊ Is the correct salt being used and is it being replenished? ◊ Is it still in circuit and not bypassed and acting as a large dead leg? ◊ Check the materials as used for the supply pipework to any
l
l High level filtration and water scavenging units (If required) ◊ Examine the maintenance records for filter changes. ◊ When filters were changed, were the seals changed or disinfected? ◊ Were the filter housings cleaned out before the new elements were fitted? ◊ Check the materials as used for the filter housings and systems. ◊ Are pressure gauges fitted to enable an indication of filter failures and blockages that may be causing problems? ◊ Are the gauges checked periodically for function and comparison readings against a known source?
Note: More scavenging systems may be required in the future
as tanked chlorinated water supplies increase
70
@ 2025 Central Sterilising Club. All rights reserved. Do Not duplicate
Made with FlippingBook. PDF to flipbook with ease
RkJPbmxpbmVTSFAK - n8MyAAAAAAA= - Njk5NzA4MTU0