Table of contents First page 79 81 Last page
Flow chart for decontamination of endoscopes
Drying and storage
Pre-clean
Manual cleaning
Automated Reprocessing EWDs must only be operated by
Leak testing
Outer surface Visually inspect Attach A/W adapter and flush all channels with low foaming neutral detergent until running clear Flush suction channel Expel excess water from channels Remove from processor Attach water resistant cap Discard biopsy port cap if breached Place in rigid container clearly marked as contaminated Transport to decontamination
Remove all detachable parts Immerse endoscope fully in water Complete leak test with complete manipulation of all angulation Deflate endoscope before
Temperature of water and volume of detergent as per manufacturer’s instructions Immerse endoscope fully in water/detergent Brush all accessible channels at least 3 times. Attach and use Irrigation devices/ injection tube sets as per manufacturers’ instructions Ensure all accessible channels come into contact with detergent Aspirate detergent from a separate container into sink Clean detachable parts, wash and brush their external surfaces Detach irrigation device Transfer to separate sink to rinse Use CE marked accessory holders
After decontamination the endoscope must be used within 3 hours unless placed in a validated endoscope drying cabinet or vacuum packed Drying cabinets should be located in a designated clean area Distal tips of endoscopes should not touch or trail on the floor of cabinets Print out parameters must be maintained for QA and tracking and traceability purposes.
appropriately trained staff Ensure all
channels are connected at the start and end of the cycle Reprocess all detachable components simultaneously with the corresponding endoscope Ensure periodic testing is carried out
commencing manual clean
as per EN standards and National guidance Ensure that machine
disinfection is carried out at the start of the day Print out parameters must be maintained
area Keep moist until
reprocessed. Best practice is within 3 hours from end of patient procedure
Manufacturers’ instructions must be followed at all times. Staff should wear full PPE in accordance with local policy
Table 7.1 Five step endoscope decontamination process
80
@ 2025 Central Sterilising Club. All rights reserved. Do Not duplicate