Table of contents First page 38 40 Last page
Risk and device classification
Reccommendation / Decontamination level
Application / procedure
• Introduced into sterile body areas; e.g. surgical instruments; biopsy forceps • In close contact with broken skin or broken mucous membrane. • In contact with mucous membranes not penetrating sterile tissue e.g. flexible endoscopes, vaginal specula. • Contaminated with
Cleaning and disinfection followed by sterilisation
High / Critical
Cleaning followed by high level disinfection or sterilisation N.B. Where high temperature steam sterilisation may damage equipment, users should consider less damaging alternative low temperature sterilisation technologies
Medium / Semi-critical
particularly virulent or readily transmissible organisms. • Before use on immunocompromised patients.
• In contact with healthy skin e.g. stethoscope; sphygmomanometer
Cleaning and low level disinfection
Low / Non critical
Table 3.1 Clinical risk and Spaulding’s classification
39
@ 2025 Central Sterilising Club. All rights reserved. Do Not duplicate
Made with FlippingBook. PDF to flipbook with ease
RkJPbmxpbmVTSFAK - n8MyAAAAAAA= - Njk5NzA4MTU0